Machine wiring
Wire numbers by machine
Select a machine or search across the complete wire register.
Wire register
328 results| Machine | Wire number | Name / definition | Optional extra | Notes | From | To | |
|---|---|---|---|---|---|---|---|
| CFXCorroflex | 0L | ACC29 ATCU local neutral | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 1 | Incoming mains cable core for L1 | NO | — | Edit | ||
| CFXCorroflex | 10 | Over-temperature thermostat O/T1 | O/T1 input | NO | — | Edit | ||
| CFXCorroflex | 100 | Z1 output / chamber power distribution | NO | — | Edit | ||
| CFXCorroflex | 1000 | Heater/blower mains distribution conductor | NO | — | Edit | ||
| CFXCorroflex | 1001 | Heater/blower mains distribution conductor | NO | — | Edit | ||
| CFXCorroflex | 1002 | Heater/blower mains distribution conductor | NO | — | Edit | ||
| CFXCorroflex | 1004 | Heater/blower auxiliary supply; also air-drying blower supply | Air drying blower supply | NO | — | Edit | ||
| CFXCorroflex | 1005 | Air saturator heater supply | NO | — | Edit | ||
| CFXCorroflex | 101 | Z1 output / chamber power distribution | Shared power/supply conductor used by ACC30 / ACC102 circuits | YES — SHAREDACC30 / ACC102 | — | Edit | ||
| CFXCorroflex | 102 | Z1 output / chamber power distribution | NO | — | Edit | ||
| CFXCorroflex | 105 | ATCU air circulation fan phase conductor | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 106 | ATCU air circulation fan phase conductor | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 107 | ATCU air circulation fan phase conductor | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 108 | ATCU compressor stage 1 phase conductor | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 109 | ATCU compressor stage 1 phase conductor | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 11 | Over-temperature thermostat O/T2 | O/T2 input | NO | — | Edit | ||
| CFXCorroflex | 110 | ATCU compressor stage 1 phase conductor | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 111 | ATCU compressor stage 2 phase conductor | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 112 | ATCU compressor stage 2 phase conductor | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 113 | Blower over-temperature thermostat O/T6 circuit | ATCU compressor stage 2 phase conductor | YES — SHAREDACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 1133 | User Programmable Digital Output 2 | YESUser programmable output 2, User Programmable Digital Output 2 | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 114 | ATCU heater contactor input phase | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 115 | Salt Spray / Water Fog pump 1-2 motor or drive interconnect | ATCU heater contactor input phase | YES — SHAREDACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 116 | Salt Spray / Water Fog pump 1-2 motor or drive interconnect | ATCU heater contactor input phase | YES — SHAREDACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 117 | Sulphur Dioxide Supply Valve | Sulphur dioxide supply valve | YESACC46 Sulphur dioxide dosing | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 1180 | User Programmable Digital Output 2 relay/contact conductor | YESUser Programmable Digital Output 2 | — | Edit | ||
| CFXCorroflex | 1181 | User Programmable Digital Output 2 relay/contact conductor | YESUser Programmable Digital Output 2 | — | Edit | ||
| CFXCorroflex | 1182 | User Programmable Digital Output 2 relay/contact conductor | YESUser Programmable Digital Output 2 | — | Edit | ||
| CFXCorroflex | 1183 | User Programmable Digital Output 2 relay/contact conductor | YESUser Programmable Digital Output 2 | — | Edit | ||
| CFXCorroflex | 1189 | Phase input to Z1 | NO | — | Edit | ||
| CFXCorroflex | 121 | Air Saturator 2 full | YESSecond air saturator option | Used when the second air saturator / door-seal control is fitted. | Edit | ||
| CFXCorroflex | 122 | ATCU condenser fan branch feed | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 123 | ACC20 waste-water trough full switch | ACC20 full switch | Common feed used by user programmable digital outputs | YES — SHAREDACC20 Waste water trough & pump, User programmable digital outputs | — | Edit | ||
| CFXCorroflex | 1233 | User Programmable Digital Output 4 | YESUser Programmable Digital Output 4, Remote controls / Two-button canopy / user outputs | — | Edit | ||
| CFXCorroflex | 124 | ATCU de-superheater fan branch feed | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 125 | ACC92 scrubber flow signal | Scrubber flow meter signal | YESACC92 Exhaust salt scrubber | — | Edit | ||
| CFXCorroflex | 126 | ATCU crankcase-heater branch feed | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 1265 | ACC29 Butterfly Valve | Refrigeration & humidity control butterfly valve | YESACC29 Refrigeration & humidity control | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 127 | ATCU crankcase heater stage 1 feed | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 128 | ATCU crankcase heater stage 2 feed | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 132 | Salt Spray / Water Fog pump 1-2 motor or drive interconnect | NO | — | Edit | ||
| CFXCorroflex | 133 | User Programmable Digital Output 1 | YESUser programmable output 1, User Programmable Digital Output 1 | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 134 | Spot Cooler | Spot cooler relay command | YESACC90 Spot cooler | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 135 | Air Saturator 2 Water/Fog changeover solenoid | YESSecond air saturator option | — | Edit | ||
| CFXCorroflex | 136 | Multi Solution changeover solenoid | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 137 | Additional Side Heater Mats | Additional side heater mats relay command | YESAdditional side heater mats | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 139 | Additional side heater mat switched conductor | YESAdditional side heater mats | — | Edit | ||
| CFXCorroflex | 14 | Moist air valve | Moist air | NO | — | Edit | ||
| CFXCorroflex | 144 | Interior illumination switched mains feed after MCB7 | NO | — | Edit | ||
| CFXCorroflex | 146 | ACC102 salt solution consumption sensor signal | Salt solution consumption sensor signal | YESACC102, ACC102 Salt solution consumption sensor | — | Edit | ||
| CFXCorroflex | 147 | Salt tank low-water alarm | Salt tank low water alarm | YESACC59 salt tank alarm option | — | Edit | ||
| CFXCorroflex | 148 | Second blower ball valve position/contact conductor | YESACC39 / ACC47 Blower ball valve | — | Edit | ||
| CFXCorroflex | 150 | ATCU stage 1 COOL solenoid switched feed | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 151 | ATCU stage 1 BYPASS solenoid switched feed | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 152 | ATCU stage 2 COOL solenoid switched feed | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 153 | ATCU stage 2 BYPASS solenoid switched feed | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 154 | ATCU DEHUMID solenoid switched feed | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 1580 | User Programmable Digital Output 4 relay/contact conductor | YESUser Programmable Digital Output 4 | — | Edit | ||
| CFXCorroflex | 1581 | User Programmable Digital Output 4 relay/contact conductor | YESUser Programmable Digital Output 4 | — | Edit | ||
| CFXCorroflex | 1582 | User Programmable Digital Output 4 relay/contact conductor | YESUser Programmable Digital Output 4 | — | Edit | ||
| CFXCorroflex | 1583 | User Programmable Digital Output 4 relay/contact conductor | YESUser Programmable Digital Output 4 | — | Edit | ||
| CFXCorroflex | 164 | ATCU OK | ATCU plant healthy | YES — SHAREDACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 17 | ATCU heater SSR1 output | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 170 | Blower ball-valve position | Blower ball valve position feedback | YESACC39 / ACC47 Blower ball valve | — | Edit | ||
| CFXCorroflex | 173 | Low water | Water low input | NO | — | Edit | ||
| CFXCorroflex | 177 | Additional Salt Spray / Water Fog pump motor/F2V interconnect | YESACC29-dependent additional spray pumps | Shown in the additional-pumps circuits with ACC29 annotations. | Edit | ||
| CFXCorroflex | 18 | Emergency stop feedback | Emergency stop input | ATCU heater SSR2 output | YES — SHAREDACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 180 | User Programmable Digital Output 1 relay/contact conductor | YESUser Programmable Digital Output 1 | — | Edit | ||
| CFXCorroflex | 181 | User Programmable Digital Output 1 relay/contact conductor | YESUser Programmable Digital Output 1 | — | Edit | ||
| CFXCorroflex | 182 | User Programmable Digital Output 1 relay/contact conductor | YESUser Programmable Digital Output 1 | — | Edit | ||
| CFXCorroflex | 183 | User Programmable Digital Output 1 relay/contact conductor | YESUser Programmable Digital Output 1 | — | Edit | ||
| CFXCorroflex | 185 | Psychometric drains | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 187 | Phase input to Z1 | NO | — | Edit | ||
| CFXCorroflex | 188 | Phase input to Z1 | NO | — | Edit | ||
| CFXCorroflex | 1889 | Phase feed from X1 to MCB30 / MCB3 | NO | — | Edit | ||
| CFXCorroflex | 189 | Phase feed after MCB30 to optional ATCU power pages | NO | — | Edit | ||
| CFXCorroflex | 1890 | ACC29 ATCU phase after MCB21 | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 19 | Canopy opened | Canopy opened input | ATCU heater SSR3 output | YES — SHAREDACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 191 | Spot cooler mains feed after MCB15 | YESACC90 Spot cooler | — | Edit | ||
| CFXCorroflex | 196 | Air Saturator 2 heater inhibit | YESSecond air saturator option | Used when the second air saturator / door-seal control is fitted. | Edit | ||
| CFXCorroflex | 198 | Heater over-temperature count / O/T10 chain | Heater over-temperature count input | NO | — | Edit | ||
| CFXCorroflex | 1981 | O/T11 thermostat intermediate conductor | NO | — | Edit | ||
| CFXCorroflex | 1L | ACC29 ATCU local phase 1 | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 2 | Incoming mains cable core for L2; also reused as AC neutral in internal circuits | Air drying blower neutral | Shared power/supply conductor used by ACC30 / ACC102 circuits | YES — SHAREDACC30 / ACC102 | — | Edit | ||
| CFXCorroflex | 20 | Auto fill valve | Auto fill | NO | — | Edit | ||
| CFXCorroflex | 21 | Air saturator full | Air saturator full input | NO | — | Edit | ||
| CFXCorroflex | 210 | Heat | ATCU HEAT input | YESACC29 Refrigeration & humidity control | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 213 | Remote PAUSE | YESRemote controls / Two-button canopy / user outputs | — | Edit | ||
| CFXCorroflex | 214 | Remote UNPAUSE | YESRemote controls / Two-button canopy / user outputs | — | Edit | ||
| CFXCorroflex | 215 | Remote START | YESRemote controls / Two-button canopy / user outputs | — | Edit | ||
| CFXCorroflex | 216 | Remote STOP | YESRemote controls / Two-button canopy / user outputs | — | Edit | ||
| CFXCorroflex | 217 | Two Button Canopy Option input | YESRemote controls / Two-button canopy / user outputs | — | Edit | ||
| CFXCorroflex | 22 | Air saturator low | Air saturator low input | NO | — | Edit | ||
| CFXCorroflex | 222 | Air Saturator 2 low | YESSecond air saturator option | Used when the second air saturator / door-seal control is fitted. | Edit | ||
| CFXCorroflex | 233 | User Programmable Digital Output 3 | YESUser Programmable Digital Output 3, Remote controls / Two-button canopy / user outputs | — | Edit | ||
| CFXCorroflex | 236 | Door/open control feed before E-stop interlock | NO | — | Edit | ||
| CFXCorroflex | 237 | Door/close control feed before E-stop interlock | NO | — | Edit | ||
| CFXCorroflex | 238 | Door seal control feed before E-stop interlock | NO | — | Edit | ||
| CFXCorroflex | 239 | Door unseal control feed before E-stop interlock | NO | — | Edit | ||
| CFXCorroflex | 24 | +24 V DC control supply | +24 V supply to optional output cards | Supply to PT100 / analogue cards including ACC34 option | Shared supply/return used by page 15 optional accessories | Shared supply/return to ACC42 / ACC46 / ACC29 accessories | Shared supply/return to ACC92 / ACC31 / ACC66 / ACC59 accessories | Shared power/supply conductor used by ACC30 / ACC102 circuits | Shared supply/return for page 19 optional accessories | Shared supply/return for ACC36 remote transmission and optional remote controls | YES — SHAREDOptional output cards, ACC34 optional PT100, ACC20 / ACC90 / additional pump options, ACC42 / ACC46 / ACC29, ACC92 / ACC31 / ACC66 / ACC59, ACC30 / ACC102, ACC39 / ACC47 / ACC05 / user outputs, ACC36 / Remote controls / Two-button canopy | Defined by the wire-identification table on page 1. ACC34 sensor leads are colour/terminal identified rather than assigned separate numeric wire IDs. | Edit | ||
| CFXCorroflex | 24L | ACC29 ATCU +24 V DC | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 25 | Open canopy | NO | — | Edit | ||
| CFXCorroflex | 250 | Air distribution bar | Air bar | NO | — | Edit | ||
| CFXCorroflex | 252 | Salt Spray Motor Run | Salt spray vertically-down motor run relay command | YESACC30 Salt spray vertically down | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 26 | Dry air valve | Dry air | NO | — | Edit | ||
| CFXCorroflex | 265 | Run | ATCU RUN input | YESACC29 Refrigeration & humidity control | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 27 | Close canopy | NO | — | Edit | ||
| CFXCorroflex | 28 | Canopy closed | Optional pneumatic sensor switch S23 connection into canopy-closed input | Canopy closed input | YES — SHAREDOptional pneumatic sensor switch S23 | The drawing labels S23 'PNEUMATIC SENSOR SWITCH OPTIONAL'. | Edit | ||
| CFXCorroflex | 286 | Humidify | ATCU HUMIDITY input | YESACC29 Refrigeration & humidity control | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 289 | P_Fill | Psychometric water fill / P_Fill | YESACC29 Refrigeration & humidity control | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 2L | ACC29 ATCU local phase 2 | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 3 | Incoming mains cable core for L3 | NO | — | Edit | ||
| CFXCorroflex | 31 | Salt Spray / Water Fog pump 1-2 motor or drive interconnect | NO | — | Edit | ||
| CFXCorroflex | 311 | Air curtain | NO | — | Edit | ||
| CFXCorroflex | 317 | ATCU heater 1 feed after protection | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 318 | ATCU heater 2 feed after protection | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 319 | ATCU heater 3 feed after protection | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 32 | Salt Spray / Water Fog pump 1-2 motor or drive interconnect | NO | — | Edit | ||
| CFXCorroflex | 323 | ATCU condenser fan switched feed | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 325 | ATCU de-superheater fan switched feed | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 33 | Blower Ball Valve(s) c/w Position Switch(es) | Blower ball valve drive | YESACC39 / ACC47 Blower ball valve | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 34 | Buzzer | NO | — | Edit | ||
| CFXCorroflex | 340 | Door seal command after emergency-stop interlock | Door seal | NO | — | Edit | ||
| CFXCorroflex | 341 | Door unseal command after emergency-stop interlock | Door unseal | NO | — | Edit | ||
| CFXCorroflex | 342 | Door sealed feedback | Door sealed | NO | Used when the second air saturator / door-seal control is fitted. | Edit | ||
| CFXCorroflex | 343 | Canopy unsealed | NO | Used when the second air saturator / door-seal control is fitted. | Edit | ||
| CFXCorroflex | 344 | ATCU compressor stage 1 contactor | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 345 | ATCU compressor stage 2 contactor | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 35-1 | Air saturator 1 heater circuit | NO | — | Edit | ||
| CFXCorroflex | 35-2 | Air saturator 2 heater circuit | NO | — | Edit | ||
| CFXCorroflex | 350 | ATCU stage 1 cool solenoid relay | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 352 | Salt spray vertically-down motor switched conductor | YESACC30 Salt spray vertically down | — | Edit | ||
| CFXCorroflex | 353 | Salt spray vertically-down motor relay output conductor | YESACC30 Salt spray vertically down | — | Edit | ||
| CFXCorroflex | 354 | ATCU dehumidification solenoid relay | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 36 | Air saturator heater relay command | Air saturator heater | NO | — | Edit | ||
| CFXCorroflex | 37 | Chamber heater relay command | Chamber heater | NO | — | Edit | ||
| CFXCorroflex | 370 | Inverter enable | NO | — | Edit | ||
| CFXCorroflex | 374 | ATCU air temperature limit (60C) | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 375 | ATCU PLC input I4 - function not named | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 376 | ATCU PLC input I5 - function not named | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 377 | ATCU PLC input I6 - function not named | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 38 | Immersion heater relay command | Immersion heater | NO | — | Edit | ||
| CFXCorroflex | 380 | ATCU recirculation fan overload | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 381 | ATCU stage 1 compressor overload | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 385 | ATCU PLC input I14 - function not named | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 387 | ATCU PLC input I16 - function not named | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 39-1 | Heater bank E1 switched conductor | NO | — | Edit | ||
| CFXCorroflex | 39-2 | Heater bank E2 switched conductor | NO | — | Edit | ||
| CFXCorroflex | 39-3 | Heater bank E3 switched conductor | NO | — | Edit | ||
| CFXCorroflex | 392 | ATCU heater SSR control | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 393 | ATCU Y36 solenoid output | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 3L | ACC29 ATCU local phase 3 | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 4 | Incoming mains cable core for N; also reused as AC live/control feed after MCB6/MCB7 | NO | — | Edit | ||
| CFXCorroflex | 401 | 8-channel analogue output to optional pump/F2V drive | YES — OPTIONAL I/OACC29 / ACC86 additional pump drives | Channel function depends on the fitted pump/option. | Edit | ||
| CFXCorroflex | 402 | Multi Solution Spray Motor M17 | YES — OPTIONAL I/OACC86 Multi Solution | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 403 | 8-channel analogue output to optional pump/F2V drive | YES — OPTIONAL I/OACC29 / ACC86 additional pump drives | Channel function depends on the fitted pump/option. | Edit | ||
| CFXCorroflex | 404 | Multi Solution Spray Motor M18 | YES — OPTIONAL I/OACC86 Multi Solution | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 405 | Salt Spray 3 / Water Fog 3 | YES — OPTIONAL I/OAdditional spray pump output | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 406 | 8-channel analogue output to optional pump/F2V drive | YES — OPTIONAL I/OACC29 / ACC86 additional pump drives | Channel function depends on the fitted pump/option. | Edit | ||
| CFXCorroflex | 407 | 8-channel analogue output to optional pump/F2V drive | YES — OPTIONAL I/OACC29 / ACC86 additional pump drives | Channel function depends on the fitted pump/option. | Edit | ||
| CFXCorroflex | 41 | ATCU PLC relay output Q2 - function not named | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 410 | Salt Spray / Water Fog pump 1-2 motor or drive interconnect | NO | — | Edit | ||
| CFXCorroflex | 412 | Salt Spray / Water Fog pump 1-2 motor or drive interconnect | NO | — | Edit | ||
| CFXCorroflex | 413 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 414 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 415 | Salt Spray / Water Fog pump 1-2 motor or drive interconnect | NO | — | Edit | ||
| CFXCorroflex | 416 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 417 | Salt Spray / Water Fog pump 1-2 motor or drive interconnect | NO | — | Edit | ||
| CFXCorroflex | 418 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 419 | Additional Salt Spray / Water Fog pump motor/F2V interconnect | YESACC29-dependent additional spray pumps | Shown in the additional-pumps circuits with ACC29 annotations. | Edit | ||
| CFXCorroflex | 42 | ATCU PLC relay output Q3 - function not named | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 422 | Additional Salt Spray / Water Fog pump motor/F2V interconnect | YESACC29-dependent additional spray pumps | Shown in the additional-pumps circuits with ACC29 annotations. | Edit | ||
| CFXCorroflex | 424 | Additional Salt Spray / Water Fog pump motor/F2V interconnect | YESACC29-dependent additional spray pumps | Shown in the additional-pumps circuits with ACC29 annotations. | Edit | ||
| CFXCorroflex | 425 | Additional Salt Spray / Water Fog pump motor/F2V interconnect | YESACC29-dependent additional spray pumps | Shown in the additional-pumps circuits with ACC29 annotations. | Edit | ||
| CFXCorroflex | 426 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 427 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 428 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 429 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 43 | ATCU air circulation fan contactor | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 430 | Comp | YES — OPTIONAL I/OACC29 Refrigeration & humidity control | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 431 | Salt Spray / Water Fog pump 3-4 motor interconnect | YESACC29-dependent additional spray pumps | Shown on circuits marked 'Fit With ACC29'. | Edit | ||
| CFXCorroflex | 432 | Salt Spray / Water Fog pump 3-4 motor interconnect | YESACC29-dependent additional spray pumps | Shown on circuits marked 'Fit With ACC29'. | Edit | ||
| CFXCorroflex | 433 | Salt Spray / Water Fog pump 3-4 motor interconnect | YESACC29-dependent additional spray pumps | Shown on circuits marked 'Fit With ACC29'. | Edit | ||
| CFXCorroflex | 434 | Salt Spray / Water Fog pump 3-4 motor interconnect | YESACC29-dependent additional spray pumps | Shown on circuits marked 'Fit With ACC29'. | Edit | ||
| CFXCorroflex | 435 | Additional Salt Spray / Water Fog pump motor/F2V interconnect | YESACC29-dependent additional spray pumps | Shown in the additional-pumps circuits with ACC29 annotations. | Edit | ||
| CFXCorroflex | 436 | Additional Salt Spray / Water Fog pump motor/F2V interconnect | YESACC29-dependent additional spray pumps | Shown in the additional-pumps circuits with ACC29 annotations. | Edit | ||
| CFXCorroflex | 437 | Additional Salt Spray / Water Fog pump motor/F2V interconnect | YESACC29-dependent additional spray pumps | Shown in the additional-pumps circuits with ACC29 annotations. | Edit | ||
| CFXCorroflex | 438 | Additional Salt Spray / Water Fog pump motor/F2V interconnect | YESACC29-dependent additional spray pumps | Shown in the additional-pumps circuits with ACC29 annotations. | Edit | ||
| CFXCorroflex | 439 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 44 | Interior illumination lamp circuit conductor | NO | — | Edit | ||
| CFXCorroflex | 440 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 441 | Salt Scrubber Valve | Exhaust salt scrubber valve | YESACC92 Exhaust salt scrubber | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 442 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 443 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 444 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 445 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 446 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 45 | Interior illumination relay command | Interior Illumination | YES — OPTIONAL I/OACC26 / Interior illumination | Named 'Interior Illumination' on the optional output card. Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 450 | Purge Valve for Salt Spray | Salt spray purge valve | YESACC30 Salt spray vertically down | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 46 | ATCU heater contactor command before OT2 | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 463 | Pressure Release Valve for Salt Spray | Salt spray pressure release valve | YESACC30 Salt spray vertically down | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 470 | GA - Fatal Error Warning | Guardian Angel fatal | YESACC05 Guardian Angel | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 471 | GA - Non-Fatal Error Warning | Guardian Angel non-fatal | YESACC05 Guardian Angel | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 472 | GA - System Healthy | Guardian Angel healthy | YESACC05 Guardian Angel | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 474 | Salt Spray 4 / Water Fog 4 | YES — OPTIONAL I/OAdditional spray pump output | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 475 | 8-channel analogue output to optional pump/F2V drive | YES — OPTIONAL I/OACC29 / ACC86 additional pump drives | Channel function depends on the fitted pump/option. | Edit | ||
| CFXCorroflex | 48 | Blower heater control / optional output | Blower Heater | YES — OPTIONAL I/OBlower heater optional output | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 480 | Spraybar parked | Spraybar parked sensor signal | YESACC31 Oscillating spray bar, ACC31/1 or ACC31/2 | — | Edit | ||
| CFXCorroflex | 4L | ACC29 ATCU PLC/control AC live | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 50 | Wash Down Valve | Wall wash down valve | YESACC42 Wall wash down | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 500 | Oscillating Spray Bar Motor Engage | Oscillating spray bar motor engage | YESACC31 Oscillating spray bar | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 502 | Oscillating spray bar motor conductor | YESACC31 Oscillating spray bar | — | Edit | ||
| CFXCorroflex | 504 | Additional Salt Spray / Water Fog pump motor/F2V interconnect | Corroflex Pump Output 3 | YESACC29-dependent additional spray pumps, Additional spray pumps | Shown in the additional-pumps circuits with ACC29 annotations. | Edit | ||
| CFXCorroflex | 51 | ATCU stage 1 bypass solenoid relay | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 519 | Air Saturator 2 / Water Fog / Salt Spray changeover solenoid circuit | Water Fog N/C solenoid / Salt Spray N/O solenoid | NO | — | Edit | ||
| CFXCorroflex | 52 | ATCU stage 2 cool solenoid relay | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 53 | ATCU stage 2 bypass solenoid relay | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 537 | Additional Salt Spray / Water Fog pump motor/F2V interconnect | YESACC29-dependent additional spray pumps | Shown in the additional-pumps circuits with ACC29 annotations. | Edit | ||
| CFXCorroflex | 538 | Additional Salt Spray / Water Fog pump motor/F2V interconnect | YESACC29-dependent additional spray pumps | Shown in the additional-pumps circuits with ACC29 annotations. | Edit | ||
| CFXCorroflex | 539 | Additional Salt Spray / Water Fog pump motor/F2V interconnect | YESACC29-dependent additional spray pumps | Shown in the additional-pumps circuits with ACC29 annotations. | Edit | ||
| CFXCorroflex | 54-1 | Immersion heater switched conductor | NO | — | Edit | ||
| CFXCorroflex | 54-2 | Immersion heater switched conductor | NO | — | Edit | ||
| CFXCorroflex | 54-3 | Immersion heater switched conductor | NO | — | Edit | ||
| CFXCorroflex | 540 | Additional Salt Spray / Water Fog pump motor/F2V interconnect | YESACC29-dependent additional spray pumps | Shown in the additional-pumps circuits with ACC29 annotations. | Edit | ||
| CFXCorroflex | 541 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 55 | ATCU PLC relay output Q13 - function not named | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 560 | De-Humidify | ATCU DEHUMID input | YESACC29 Refrigeration & humidity control | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 563 | 8-channel analogue output to optional pump/F2V drive | YES — OPTIONAL I/OACC29 / ACC86 additional pump drives | Channel function depends on the fitted pump/option. | Edit | ||
| CFXCorroflex | 564 | 8-channel analogue output to optional pump/F2V drive | YES — OPTIONAL I/OACC29 / ACC86 additional pump drives | Channel function depends on the fitted pump/option. | Edit | ||
| CFXCorroflex | 57 | ATCU heater contactor command after OT2 | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 58 | Salt Spray Air Valve | Salt Spray Air Solenoid | YESAdditional spray pumps | Named on the sheet titled 'OPTIONAL OUTPUTS'. Optional if additional pumps are fitted. | Edit | ||
| CFXCorroflex | 580 | User Programmable Digital Output 3 relay/contact conductor | YESUser Programmable Digital Output 3 | — | Edit | ||
| CFXCorroflex | 581 | User Programmable Digital Output 3 relay/contact conductor | YESUser Programmable Digital Output 3 | — | Edit | ||
| CFXCorroflex | 582 | User Programmable Digital Output 3 relay/contact conductor | YESUser Programmable Digital Output 3 | — | Edit | ||
| CFXCorroflex | 583 | User Programmable Digital Output 3 relay/contact conductor | YESUser Programmable Digital Output 3 | — | Edit | ||
| CFXCorroflex | 59 | Base fill valve | Base fill | NO | — | Edit | ||
| CFXCorroflex | 598 | ACC86 Multi Solution pump / F2V / motor interconnect | Multi Solution Spray Motor M19 | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 60 | High Water Warning Lamp | High water lamp | YESACC66 High water lamp | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 601 | ACC86 Multi Solution pump / F2V / motor interconnect | Multi Solution Spray Motor M20 | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 604 | ACC86 Multi Solution pump / F2V / motor interconnect | Multi Solution Spray Motor M21 | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 608 | ACC86 Multi Solution pump / F2V / motor interconnect | Multi Solution Spray Motor M22 | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 612 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 614 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 615 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 616 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 617 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 618 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 619 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 620 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 63 | Over-temperature thermostat O/T8 | O/T8 input | NO | — | Edit | ||
| CFXCorroflex | 637 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 638 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 639 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 64 | Over-temperature thermostat O/T7 | O/T7 input | NO | — | Edit | ||
| CFXCorroflex | 640 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 641 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 642 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 643 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 644 | ACC86 Multi Solution pump / F2V / motor interconnect | YESACC86 Multi Solution | — | Edit | ||
| CFXCorroflex | 650 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 651 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 652 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 653 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 654 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 655 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 656 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 657 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 658 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 659 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 660 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 661 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 662 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 663 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 668 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 669 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 670 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 671 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 672 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 673 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 674 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 675 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 676 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 677 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 678 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 679 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 68 | Cool | ATCU COOL input | YESACC29 Refrigeration & humidity control | Named on the sheet titled 'OPTIONAL OUTPUTS'. | Edit | ||
| CFXCorroflex | 680 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 681 | ECC1 additional atomiser pump / F2V / motor interconnect | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 686 | ECC1 Pump 1 | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 687 | ECC1 Pump 2 | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 688 | ECC1 Pump 3 | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 689 | ECC1 Pump 4 | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 690 | ECC1 Pump 5 | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 691 | ECC1 Pump 6 | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 692 | ECC1 Pump 7 | YESACC25 ECC1 additional atomisers | — | Edit | ||
| CFXCorroflex | 74 | Salt Spray / Water Fog pump 1-2 motor or drive interconnect | NO | — | Edit | ||
| CFXCorroflex | 75 | Salt Spray / Water Fog pump 1-2 motor or drive interconnect | NO | — | Edit | ||
| CFXCorroflex | 76 | Pump 1 pulse | NO | — | Edit | ||
| CFXCorroflex | 77 | Pump 2 pulse | NO | — | Edit | ||
| CFXCorroflex | 78 | ATCU PLC input I7 - function not named | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 79 | ATCU heater contactor C4 auxiliary feedback | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 8 | 0 V DC common | 0 V return for optional output cards | Supply to PT100 / analogue cards including ACC34 option | Shared supply/return used by page 15 optional accessories | Shared supply/return to ACC42 / ACC46 / ACC29 accessories | Shared supply/return to ACC92 / ACC31 / ACC66 / ACC59 accessories | Shared power/supply conductor used by ACC30 / ACC102 circuits | Shared supply/return for page 19 optional accessories | Shared supply/return for ACC36 remote transmission and optional remote controls | YES — SHAREDOptional output cards, ACC34 optional PT100, ACC20 / ACC90 / additional pump options, ACC42 / ACC46 / ACC29, ACC92 / ACC31 / ACC66 / ACC59, ACC30 / ACC102, ACC39 / ACC47 / ACC05 / user outputs, ACC36 / Remote controls / Two-button canopy | Defined by the wire-identification table on page 1. ACC34 sensor leads are colour/terminal identified rather than assigned separate numeric wire IDs. | Edit | ||
| CFXCorroflex | 80 | Blower 1 tacho | NO | — | Edit | ||
| CFXCorroflex | 81 | Blower 2 tacho | NO | — | Edit | ||
| CFXCorroflex | 83 | ATCU stage 1 HP/LP signal contact | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 84 | ATCU stage 1 HP signal | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 85 | Base drain valve | Base drain | NO | — | Edit | ||
| CFXCorroflex | 86 | ATCU PLC input I15 - function not named | YESACC29 Refrigeration & humidity control | — | Edit | ||
| CFXCorroflex | 87 | Phase feed after MCB30 to optional ATCU power pages | NO | — | Edit | ||
| CFXCorroflex | 870 | ACC29 ATCU phase after MCB21 | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 88 | Phase feed after MCB30 to optional ATCU power pages | NO | — | Edit | ||
| CFXCorroflex | 880 | ACC29 ATCU phase after MCB21 | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 887 | Phase feed from X1 to MCB30 / MCB1 | NO | — | Edit | ||
| CFXCorroflex | 888 | Phase feed from X1 to MCB30 / MCB2 | NO | — | Edit | ||
| CFXCorroflex | 8L | ACC29 ATCU 0 V DC | YESACC29 Refrigeration & humidity control | Pages 22 and 24 are titled ACC29 POWER MANAGEMENT (-20 / -40). | Edit | ||
| CFXCorroflex | 91 | Low air pressure | Low-air input | NO | — | Edit | ||
| CFXCorroflex | 96 | Air saturator heater inhibit | Air saturator inhibit input | NO | — | Edit | ||
| CFXCorroflex | 97-1 | Blower E6 circuit conductor | NO | — | Edit | ||
| CFXCorroflex | 97-2 | Blower E7 circuit conductor | NO | — | Edit | ||
| CFXCorroflex | 98 | Blower over-temperature thermostat O/T3 circuit | NO | — | Edit | ||
| CFXCorroflex | E | Supply earth / protective earth | NO | Defined by the wire-identification table on page 1. | Edit | ||
| CFXCorroflex | L1 | Mains supply phase L1 | NO | Defined by the wire-identification table on page 1. | Edit | ||
| CFXCorroflex | L2 | Mains supply phase L2 | NO | Defined by the wire-identification table on page 1. | Edit | ||
| CFXCorroflex | L3 | Mains supply phase L3 | NO | Defined by the wire-identification table on page 1. | Edit | ||
| CFXCorroflex | N | Mains neutral | NO | Defined by the wire-identification table on page 1. | Edit | ||
| CFXCorroflex | N1 | Neutral before RCD / X1 | NO | — | Edit | ||
| CFXCorroflex | N2 | Neutral after RCD / X1 distribution | NO | — | Edit |